    Fluke 1658205 9130 Series Portable Infrared Calibrator

    By Fluke | Mfr#: 1658205 | Tiger#: TS53712
    Fluke 9130 Series Portable Infrared Calibrator FLU1658205-
    • Fluke 9130 Series Portable Infrared Calibrator FLU1658205-
    • Fluke 9130 Series Portable Infrared Calibrator FLU1658205-
    • Fluke 9130 Series Portable Infrared Calibrator FLU1658205-
    • Fluke 9133-156 Portable Infrared Calibrator with Solid-State Cooling Technology FLU1658222

    Highlights

    • Choice of 2 models
    • Used to calibrate non-contact infrared thermometers from -22 to 932 °F
    • Large field of view area for visual difference in thermometers
    • Temperature-regulated blackbody target
    • Target size: 2.25 inches
    • Temperature scan rate control
    • Adjustable readout of either Celsius or Fahrenheit
    • RS-232 interface with software
    More Product Details What's Included
    $5842.03
    • Product Options
      Select Model

    Ships Out Within: 10 - 14 Business Days

    Overview for Fluke 1658205 9130 Series Portable Infrared Calibrator

    The Fluke 9130 Series Portable Infrared Calibrator is a precision-tool that is used to calibrate non-contact infrared (IR) thermometers from -22 to 932 °F. This IR calibrator gives a large field of view area for visual differences in many infrared thermometers. Also, it has a large, temperature-regulated blackbody target that has a diameter of 2.25 inches.

    This portable calibrator features a temperature scan rate control which uses a precision platinum RTD as a sensor. The technician can also choose Celsius or Fahrenheit unit with its adjustable readout feature. Plus, the 9130 series has an eight set-point memory and an RS-232 interface with software for an easier calibration experience. In addition, this Fluke IR calibrator features a small and compact design with an LED display for easy data readout.

    Configuration Options

    The Fluke 9130 Series Portable Infrared Calibrator has two available models with a solid-state cooling technology. Please choose from the drop-down menu above.

    • 1658205: Fluke 9132-156 Portable Infrared Calibrator
    • 1658222: Fluke 9133-156 Portable Infrared Calibrator with Solid-State Cooling Technology


    Model #


    Fluke 9133


    Fluke 9132


    Temperature Range


    50 °C to 500 °C (122°F to 932°F)


    –30 °C to 150 °C at 23 °C ambient
    (–22 °F to 302 °F at 73 °F ambient)


    Accuracy


    ±0.9 °F at 212 °F (±0.5 °C at 100 °C)
    ±1.4 °F at 932 °F (±0.8 °C at 500 °C)


    ±0.72 °F (0.4 °C)


    Stability


    ±0.18 °F at 212 °F (±0.1 °C at 100 °C)
    ±0.54 °F at 932 °F (±0.3 °C to 500 °C)


    ±0.18 °F (±0.1 °C)


    Target Size


    2.25" (57 mm)


    2.25" (57 mm)


    Target Emissivity


    0.95 (± 0.02 from 8 to 14 μm)


    0.95 (± 0.02 from 8 to 14 μm)


    Resolution


    0.1 °


    0.1 °


    Heat Time


    30 minutes (50 °C to 500 °C)


    15 minutes (25 °C to 150 °C)


    Cooling Time


    30 minutes (500 °C to 100 °C)


    15 minutes (25 °C to -20 °C)


    Computer Interface


    RS-232 I/O included with 9930 Interface-it software


    RS-232 I/O included with 9930 Interface-it software


    Power


    115 V ac (±10 %), 3 A or
    230 V ac (±10 %), 1.5 A, switchable, 50/60 Hz, 340 W


    115 V ac (±10 %), 1.5 A, or
    230 V ac (±10 %), 1.0 A, switchable, 50/60 Hz, 200 W


    Size (H x W x D)


    6" H x 11.25" W x 10.5" D (152 x 286 x 267 mm)


    4" H x 6" W x 7" D (102 x 152 x 178 mm)


    Weight


    10 lbs (4.6 Kg)


    4 lbs (1.8 Kg)


    NIST-Traceable Contract Calibration


    Data at 50, 100, 200, 250, 300, 400 and 500 °C


    Data at -30, 0, 25, 75, 100, 125 and 150 °C